DanioVision - 6 Arena Settings

Last updated: Jul 25, 2026

6 Arena Settings

Define The Arenas

Choose Setup > Arena Settings > Arena Settings 1.

To define the arenas automatically

Follow these instructions if your experiment was created for well plates with 6, 12, 24, 48 or 96 round wells, or 96 square wells, and this is

the first new Arena Settings in your experiment. In other cases, go to To

define the arenas manually on page 52.

tip Use well plates with no notches between wells. These result in

better well detection. Also make sure that the borders of all wells are within the camera image. 1. In the Grab Background Image window, click the Grab button to grab a background image of an empty well-plate from the camera image. 2. If EthoVision XT can detect the wells, the outline of the arenas are displayed over the background image. A message appears asking whether the arenas overlap with the wells in the image. 3. Do one of the following: If there is good overlap like in the picture below, click Yes. If needed, you can always make small adjustments using the Arrange Arenas function (see page 53). If the arenas clearly do not overlap with the wells, or arenas

overlap with each other, click No. The Arrange Arenas window

appears automatically. See arrange the arenas on page 53 for

details. 4. Click the 1. Draw Scale to Calibrate button. Draw a line over the entire well-plate. Next, enter its size. Do this horizontally as well as vertically. 5. Click Validate Setup to validate the Arena Settings. Common errors are overlapping arenas, or zones falling outside an arena. If the setup is valid, continue with the instructions below.

Notes

Arenas are marked with labels A1, A2, ..., B1, B2, ... according to your

well plate (letters indicate rows; numbers indicate columns) Automatic detection of the wells in the video image has been tested with the DVOC-0040 and DVOC-0041. Success of automatic detection much depends on image contrast, focus, and distortion, and therefore we cannot guarantee it will work in all situations and with video made with previous DanioVision versions. Automatic detection also does not work if water is not present in the well plate. If automatic detection of the wells does not work, it could be due to the well plate being not centered in the video image. See Fine camera adjustments on page 122. When the well plate image is optimal and centered, and still

EthoVision XT shows the message Well plate detection failed, click

OK and proceed with To define the arenas manually on page 52.

To start automatic well plate detection, click the Well plate

detection button

at the bottom of the Arenas and Zones

section of the Arena Settings pane.

To define the arenas manually

The following instructions apply when the automatic well plate detection does not work properly. In that case you must define one arena and then make duplicates. If you used a predefined template to set up your experiments, go to arrange the arenas on page 53. 1. Click the 1. Draw Scale to Calibrate button, draw a line over the entire well-plate and enter its size. Do this horizontally as well as vertically. 2. If the current arenas are of shape different from that of the wells, select all arenas and press Delete. 3. Click 2. Select Shape and Draw Arena and use one of the shapes to create the first, top-left, arena. Make sure the label of the first arena (for example A1) points to the inside of the drawn arena. 4. Click the outline of the arena or its label and then click the 4. Copy Arena button.

In the window that appears select All other arenas. Select the

number of rows and click OK. 5. Make sure the predefined arenas cover the wells. If you need to

adjust the arena size and position, see arrange the arenas on page

53.

Arrange The Arenas

  1. in the Arena Settings window, click the 5. Arrange Arenas button.

In the Arrange Arenas window, do the following:

  1. To move all arenas: Under 1. Move Grid To move up/down: Click the first slider. Move it the right (to move the arenas down) or left (to the arenas move up). To move to the right/left: Click the second slider. Move it the right (to move arenas to the right) or left (to move them to the left). You can also click the up/down arrow buttons of the corresponding box or enter a number in the box.
  2. To scale all arenas: Under 2. Scale Arenas Resize all shapes by using the sliders, by pressing the up/down arrows of the corresponding box or by entering a number in a box. To increase the height of all arenas, use the first slider. To increase the width of all arenas, use the second slider

By default the option Keep proportions is not selected. Then you

can resize the height and width of all shapes separately. If you

select Keep proportions, you can change either the height, or the

width while keeping the aspect ratio. 4. To rotate all arenas: Under 3. Rotate Grid Optionally, rotate all shapes. This is only necessary if the well-plate is not positioned straight. The shapes rotate around the center of gravity of Arena 1. 5. To change the space between arenas: Under 4. Arena Spacing Change the space between the shapes.

By default, the option Keep proportions is not selected. Then you

can change the space between the shapes separately in the vertical and horizontal direction (relative to the first row and first column,

respectively). If you select Keep proportions, change the spacing in

either the horizontal or vertical direction while keeping the aspect ratio.

tip First position the upper left arena in the desired position and

make desired size (use scale). Next adjust arena spacing so all

arenas are correctly positioned. See Using the Multiple Arena Setup

in the EthoVision XT Help for information on setting up multiple arenas. 6. Click Validate Setup to validate the Arena Settings. Common errors are overlapping arenas, or zones falling outside an arena. If the setup is valid, continue with the instructions below.

To edit a single arena

Click the row in the Arena Settings pane for that arena. The corresponding arena is highlighted.

For details on editing shapes, in the EthoVision XT Help see Arena

Settings > Work with shapes.

Assign The Stimuli To The Arenas

Follow this procedure if you want to apply stimuli during a trial, for example a light pulse or a tapping stimulus.

important If you created an experiment with a pre-defined template

and selected the correct DanioVision arena template, the White Light, or Tapping Stimulus has been assigned automatically to the arenas. The steps below are then not necessary. 1. At the bottom of the Arena Settings pane, click the Arena - Hardware Mapping button. 2. In the Arena - Hardware Mapping window, do one or both depending on what you want to use: Click Add device. Under Arena 1, select White Light Stimulus. Click Add device. Under Arena 1, select Tapping Stimulus.

The devices White Light Stimulus and Tapping Stimulus only need

to be assigned to Arena 1. They are in fact applied to all arenas. 3. Click OK.

For more information

See also Test the hardware devices in the Reference Manual - Trial

and Hardware Control in EthoVision XT.

To set up the Tapping Device, see also The DanioVision Tapping

Device on page 72.