DanioVision TCU - Feedback And Error Messages
Last updated: Jul 30, 2026
Feedback and Error Messages
The DanioVision Temperature Control Unit gives feedback about its status with four LED lights and error codes on the temperature display.
LED Indicators
| Status | LED | Explanation | User Action |
|---|---|---|---|
| Filling / Draining | Orange – blinking | System is pumping. |
|
| Filling / Draining | Red – continuous | Time out. There has been no flow in the water circulation and/or the cooling liquid circulation for more than 10 minutes. | Reset the system by turning the Operation switch to Off and then back to Fill/Drain. |
| Heating / Cooling | Green – blinking | The water is being heated/cooled to the set temperature. | Wait. No action needed. |
| Ready | Green – continuous | The set temperature has been reached and the system is ready to be used. | The Temperature Control Unit is ready to be used and an experiment can be started. |
| Error | Red – 1 blink – repeatedly | Insufficient flow in the water circulation. |
Look for the source of the problem. This may be:
The Fill/Drain mode can be used to localize the problem. Reset the system by setting the Operation switch to Off and then to Fill/Drain. |
| Error | Red – 2 blinks – repeatedly | Insufficient flow in the cooling liquid circulation. |
Look for the source of the problem. This may be:
The Fill/Drain mode can be used to localize the problem. Reset the system by setting the Operation switch to Off and then to Fill/Drain. |
| Error | Red – 3 blinks – repeatedly | The fan is not spinning, or not spinning at sufficient speed. | Check if there is anything blocking the fan. If this is not the case, please contact Noldus support. |
| Error | Red – 4 blinks – repeatedly | System is overheated. |
Turn off the system and end the experiment. Cooling liquid can be cooled down by switching the Operation switch to Fill/Drain. Search for the possible cause of the overheating:
|
| Error | Red – 5 blinks – repeatedly | Heating and cooling have been activated simultaneously. An electrical problem has occurred. | Please contact Noldus support. |
| Error | Red – 6 blinks – repeatedly | A problem with the data logger has occurred. |
|
| Error | Red – continuous | A temperature deviation larger than 0.5°C above or below the set temperature has been detected. Water conditioning will continue to try to resolve the situation. Once resolved, the status of the system returns to 'Ready'. | Depending on your protocol you can choose to continue or to abort the experiment. |
Temperature Display
Status Codes
- The water is being heated to the set temperature.
- The water is being cooled to the set temperature.
- The current water temperature is at least 0.5°C below the set temperature. Shown together with temperature alarm code
E05. The red Error LED (continuous) is on. - The current water temperature is at least 0.5°C above the set temperature. Shown together with temperature alarm code
E04. The red Error LED (continuous) is on. - Error mode indicator.
- This is a system setting indicator, which blinks when the TCU is in heating/cooling mode. No action need be taken.
- LCD is off: A possible reason is that one of the flat cables coming from the LCD display is not properly connected. Open the DV-TCU, locate the multi-color flat cable (3rd to the right of the SD card), disconnect it and connect it again.
Error Codes
| Error Code | Explanation | User Action |
|---|---|---|
E01 |
The temperature sensor is not detected. | Please connect the temperature sensor cable (with cinch-plugs on both ends). |
E04 |
'Too high' temperature alarm. The measured temperature is 0.5°C higher than the set temperature. | Depending on your protocol you can choose to continue or abort the experiment. Note that this error also occurs while the system is heating/cooling before the start of the experiment; the system continues to heat/cool to try to resolve the error. |
E05 |
'Too low' temperature alarm. The measured temperature is 0.5°C lower than the set temperature. | Depending on your protocol you can choose to continue or abort the experiment. Note that this error also occurs while the system is heating/cooling before the start of the experiment. |
Ed1 |
System error. | See the Error status entry in the LED Indicators section above. |
Source: DanioVision TCU Reference Manual, Noldus Information Technology