EthoVision XT 19 - Manually scored behavior
Last updated: Jul 29, 2026
Definition
Behaviors defined in the Manual Scoring Settings are analyzed as dependent variables with discrete states, like In zone, Movement etc.
How to Specify a Manually Scored Behavior
Notes
- Not [behavior name]: For example, if you define Rearing as a Start-Stop behavior and score it one time during the test (not at the start of the trial), then the results will be:
- Frequency of Rearing = 1.
- Frequency of Not Rearing = 2. The first occurrence of Not Rearing is at the start of the trial, and the second occurrence is after you press the Stop code for Rearing.
See Also
- Export manually-scored behaviors
Hardware and Advanced
- Trial Control event
- Trial Control state
- Trial Control variable
- External data
- Free interval
- Multi condition
- JavaScript continuous
- JavaScript event
- JavaScript state
See Also
For Hardware variables related to devices like the Pellet dispenser, the Lickometer, PhenoWheel or the PhenoTyper Top Unit, see the Analysis of Trial Control data in the EthoVision XT 19 - Trial and Hardware Control - Reference Manual. To open this manual, choose Apps > Noldus > EthoVision XT 19 Other Documentation.
Definition
A point event, with no duration, defined by an element of Trial and Hardware Control (condition, action, rule/sub-rule and reference). A Trial Control event has no duration. For Trial Control events you can only calculate frequency and latency of first or last occurrence.
How to Specify Trial Control Event
Notes
- Statistics of Cumulative duration and Latency can only be a multiple of the sample interval (= 1/sample rate). For example, when you define a condition Subject in zone A for >= 3 s, this condition is met when the time elapsed from its activation exceeds 3 s. If the sample rate is 12.5 frames/s (thus the sample interval is 1/12.5 = 0.08 s), the condition is met at the first multiple of 0.08 greater than 3 s, that is 3.04 s. This affects data analysis; for example, the latency of the event Condition becomes true is 3.04 s.
- For Trial Control events based on Conditions, the sample time at which a Trial Control event is scored also depends on the statistic used in the condition:
- If you use Current to define the condition (for example:
when Current In zone = true), the Trial Control event is scored at the expected sample time. For example, when the animal enters the zone (and thusCurrent = truefor that sample). See also A in the figure in Trial Control state, where the start of the Trial Control state is like a Trial Control event. - If you use any other statistic to define the condition (for example,
when Frequency of In zone >= 1), consider the following scenarios:- When the condition becomes true after the condition box is activated, the Trial Control event is scored at the expected sample time. See How the trial control instructions are executed for an explanation of the terms true and active.
- When the condition is already true when the condition box is activated (for example, a condition
Frequency of In zone = 1is activated when the animal is already in the zone), such a statistic is only evaluated at the next sample (or in the second next sample, in the case of Heading). In that case, the Trial Control event is scored one sample (or two) later than expected from the condition (see B in the figure in Trial Control state). However, for the consecutive frequencies ofIn zone = 2, 3, etc., the condition is already active by definition; therefore the Trial Control event is scored at the expected time, when the condition becomes true.
- If you use Current to define the condition (for example:
See Also
- The EthoVision XT 19 - Trial and Hardware Control - Reference Manual. To open this manual, choose Apps > Noldus > EthoVision XT 19 Other Documentation.
Application
You can use Trial Control event to test whether trial control works as expected. For example, visualize the Trial Control event condition true and plot it together with the video to check that the condition is met at the correct time. You can also define a Trial Control event to calculate the number of occurrences of specific actions, such as dropping a food pellet in a learning experiment.
Source: EthoVision XT 19 Help, Noldus Information Technology