CatWalk XT – 'Created With a Newer Version' Error When Opening an Experiment on an Analysis Computer
Last updated: Nov 26, 2026
Applies to: CatWalk XT 10.6 (analysis computer) opening experiments from 10.7
Problem
When you open a CatWalk XT experiment on an analysis computer, you see: 'Experiment could not be opened; it was created with a newer version of CatWalk XT.'
Details
The error appears when the data was acquired on a computer with a newer version of CatWalk XT (10.7) and the analysis computer has an older one (10.6). An older version cannot open an experiment from a newer version, and once an experiment is opened in a newer version it is upgraded. If the newer version was never used on the experiment, a corrupt version.dat file in the experiment folder can cause the same message (see the published articles 'Experiment fails to open – corrupted version.dat file' and 'My CatWalk XT experiment will not open as it reports that it was made with a previous (or newer) version'). Reported with CatWalk XT 10.6 on the analysis computer and 10.7.712 on the acquisition computer.
Solution
- On the acquisition computer, check the version with Help > About CatWalk XT.
- Check the version on the analysis computer in the same way.
- If the analysis computer has an older version, install on it the same version as, or a newer version than, the acquisition computer. Contact Noldus Support if you need the installer for that version.
- If both computers have the same version, or the newer version was never used on the experiment, the version.dat file may be corrupt. Follow the published version.dat articles.
- To keep a copy in the old format, duplicate the experiment folder before you open it in a newer version.