Basler GigE Camera 'Device Unreachable' Error – Subnet Mismatch When Adding a Second Camera to a Multi-Port Network Card

Last updated: Nov 23, 2026

Problem

When connecting a second (or additional) Basler GigE camera to a different port on your multi-port GigE network card and assigning it a static IP address in Pylon IP Configurator, you receive the error 'The device is unreachable! Currently you will not be able to use the device.'

Details

Each physical port on a multi-port PCIe GigE network adapter is a separate network interface with its own IP address and subnet. If you assign a camera an IP address that is on the same subnet as a camera already connected to a different port (e.g., both cameras on the 192.168.200.x subnet but plugged into different ports), the network card cannot route traffic correctly and the second camera appears unreachable. For example, if your first port is configured as 192.168.100.1 and your second port as 192.168.200.1, a camera on port 2 must use an address in the 192.168.200.x range, not the 192.168.100.x range used by port 1. This applies to EthoVision XT with Basler GigE cameras on any multi-port network adapter.

Solution

1. **Identify the IP address of each network adapter port.**
- Open **Windows Settings > Network & Internet > Change adapter options** (or open the **Network Connections** control panel).
- Find the network adapters corresponding to each port on your GigE card. Note the IP address assigned to each port (e.g., port 1 = 192.168.100.1, port 2 = 192.168.200.1, port 3 = 192.168.150.1).
2. **Determine the subnet for the port the new camera is connected to.** The camera must be assigned an IP address on the same subnet as the adapter port it is physically connected to. If port 3 on your card is 192.168.150.1 with subnet mask 255.255.255.0, then the camera on port 3 must have an address in the range 192.168.150.2–192.168.150.254.
3. **Assign a unique IP address on the correct subnet.**
- Open **Pylon IP Configurator**.
- Select the camera connected to port 3.
- Set the IP address to match the subnet of that port (e.g., 192.168.150.2), with the same subnet mask (255.255.255.0) and the gateway set to the adapter's own IP (192.168.150.1) or left blank.
- Apply the change.
4. **Confirm each camera has a unique address.** No two cameras or adapters should share the same IP address. Use a different third octet for each port's subnet if possible (e.g., 192.168.100.x for port 1, 192.168.200.x for port 2, 192.168.150.x for port 3).
5. **Verify detection in Pylon Viewer.** Open Pylon Viewer and confirm all cameras appear and can be opened individually.
6. **Verify detection in EthoVision.** Open EthoVision XT, go to **Acquisition > Select video source**, and confirm all cameras are listed and selectable.
7. **If cameras are still not detected**, reinstall Pylon and confirm the Basler DirectShow driver is installed. See the articles *Basler GigE Camera Not Detected by EthoVision or Pylon Viewer* and *Basler GigE Camera IP Configuration – Static IP Recommended Over DHCP for Reliable Detection* for additional steps.

If you continue to see the 'device unreachable' error after following these steps, open a support request at https://noldus.com/app/home.

shopping_bag
check_circle

Thank you!

We'll get back to you shortly.

error

Please correct the following errors:

error

error

error

error

By clicking Submit, you consent to Noldus processing your data as described in our privacy policy.

check_circle

Thank you for registering!

We look forward to seeing you at the NoldusViso launch event.

error

Please correct the following errors:

error

error

error

error

error

By clicking Register, you consent to Noldus processing your data as described in our privacy policy.