menu_book

Knowledge Base

Documentation, guides, and resources for Noldus products.

DanioVision - D System specifications

Last updated: Jul 25, 2026

D - System specifications

System Specifications

Daniovision Dvoc-0041

Power supply

Power requirements: 24 Volt DC, 2.0Amp (max) Mains adapter: input 100-240 Volt AC, output 24 Volt DC, 3.0Amp

Camera

Brand, type: Basler acA1300-60gm Interface: Gigabit Ethernet (RJ45 connector, CAT5e or higher) Sensor: 1/1.8" Progressive Scan CMOS, monochrome Max. resolution: 1280 x 1024 Max. frame rate: 60 fps

Lens

12 mm Megapixel, C-mount, F1.4, includes IR pass filter Optional lens: 8 mm Megapixel, C-mount, F1.4, includes IR pass filter

Video performance (H x V pixels)

Default: 1280 x 960 at 30 fps (in combination with 12 mm lens) Optional: 640 x 480 at 60 fps (in combination with 12 mm lens) 800 x 600 at 60 fps (in combination with 8 mm lens) Maximum frame rates as measured in 96 well plate setup.

Backlight unit

Combined infrared (IR) and white light array Infrared wavelength (typical): 950 nm D - System specifications According to published research, zebrafish are sensitive to infrared light, though not above 910 nm. See Shcherbakov et al. 2013, PLoS

ONE 8(5): e64429. The IR light provided by DanioVision is therefore

unlikely to affect the fish's behavior. White light color temperature (typical): 5500ºK White light level controllable via USB (EthoVision XT). Intensity range: 0 to ±10000 lux (light level measured directly at the bottom of the water basin). With the standard White Light settings, the light intensities of the white light are approximately:

note If you set the Low Light switch to L (low), the highest light

level (100%) is around 20 lux. See configure the daniovision

white light on page 36 for more information.

note Light intensity in DVOC-0041 can be set higher than in DVOC-

  1. Here the reference values if you have DVOC-0040: You may want to replicate light conditions that you set in DanioVision DVOC-0040 in a new experiment with DanioVision DVOC-0041. For example, if you set light intensity to 50% in

Percentage

Steps

Intensity (lux)

1000 2500 2100 5300 3000 8000 4095 10500

Percentage

Steps

Intensity (lux)

1% 25% 1025 1450 50% 2050 2950 75% 3000 4400 100% 4096 5450 D - System specifications DanioVision DVOC-0040, this gave 2950 lux. The percentage to be set in DanioVision DVOC-0041 is (2950/5300)*50 = approx. 28%.

Multi basin

Water inlet, water outlet and overflow Basin size 158 x 132 mm Supports ANSI, SBS compatible micro plates (L =127.76± 0.5 mm (5 1/32" ± 1/64"), W=85.48 ± 0.5 mm (3 23/64" ± 1/64") with maximum height of 27mm (1 1/16") Supports Petri dishes up to 90 mm in diameter Temperature sensor supporting the use of Noldus DVTCU (Temperature Control Unit)

Water connections

Water inlet/outlet: tube fitting 8 mm Water overflow: tube fitting 10 mm

Tubes included

1.5 m Clear PVC Hose, 8 mm OD, 5 mm ID for water inlet (green color marked) 1.5 m Clear PVC Hose, 8 mm OD, 5 mm ID for water outlet (yellow color marked) 1.5 m Clear PVC Hose, 10 mm OD, 8 mm ID for water overflow (red color marked)

PC interface

USB-2 compatible

External I/O

TTL input/output (4x modular RJ45 connector): 4 x 2 TTL input 4 x 2 TTL output D - System specifications TTL expansion port 1: 4 x TTL output (open collector output) 1 x input camera 1 x output camera TTL expansion port 2 (1x modular RJ45 connector:

System start-up time (cold start)

~ 15 minutes

Dimensions

61 x 32 x 46 cm (24 1/64" x 12 19/32" x 18 7/64") (h x w x d) in closed position Door opening 36 x 29 cm (14 11/64" x 11 27/64") (h x w)

Weight

15 kg excluding power supply, external tubes and cables

Cleaning the DanioVision Observation Chamber

We recommend to use a propriety glass-cleaning fluid to clean the outside of the DanioVision Observation Chamber.

Declaration of conformity

See page 136. D - System specifications

Daniovision Tapping Device

Power

Power requirements: 24 Volt DC, 0.4Amp (max)

Features

Controlled from EthoVision XT 10.1 or later with selectable stimulus intensity (1-8) Delivers a discrete 'tap' to the well plate holder Maximum tapping rate of ±3 taps per second

Connectors

Molex 2 pin - Power (24V) Molex 4 pin - From Tapping Device to control box Multi connector (blue) - For TTL control signal

Dimensions

Tapping Device (including aluminum bracket): 6 x 24 x 7.5 cm (2 23/64" x 9 29/64" x 2 61/64") (h x w x d) Control box: 2.5 x 10 x 6.5 cm (63/64" x 3 15/16" x 2 9/16") (h x w x d)

Weight

~ 300g (Tapping Device, bracket, control box) If not pre-assembled in the DVOC-0041, the package includes mounting material

Compatible DanioVision Chambers

DVOC-0040, DVOC-0041 D - System specifications

Daniovision Toplight Unit

White light

30.000 lux at 25cm distance (well plate), 1584 lm Color temperature 4000K Array of 408 LED's Max. voltage 18 Volt

Red light

192 lm, wavelength 623 nm Array of 252 RGB LED's Max. voltage 24 Volt

Green light

384 lm, wavelength 525 nm Array of 252 RGB LED's Max. voltage 24 Volt

Blue light

240 lm, wavelength 470 nm Array of 252 RGB LED's Max. voltage 24 Volt

Unit Dimensions

270 x 220 x 30mm (10 5/8" x 8 21/32" x 1 3/16") D - System specifications

Declaration Of Conformity

D - System specifications


Source: Reference Manual - DanioVision DVOC-0041

shopping_bag
check_circle

Thank you!

We'll get back to you shortly.

error

Please correct the following errors:

error

error

error

error

By clicking Submit, you consent to Noldus processing your data as described in our privacy policy.