System Specifications

DanioVision DVOC-0041

Power Supply

  • Power requirements: 24 Volt DC, 2.0 Amp (max)
  • Mains adapter: Input 100–240 Volt AC, output 24 Volt DC, 3.0 Amp

Camera

  • Brand, type: Basler acA1300-60gm
  • Interface: Gigabit Ethernet (RJ45 connector, CAT5e or higher)
  • Sensor: 1/1.8" Progressive Scan CMOS, monochrome
  • Max. resolution: 1280 x 1024
  • Max. frame rate: 60 fps

Lens

  • Standard lens: 12 mm Megapixel, C-mount, F1.4, includes IR pass filter
  • Optional lens: 8 mm Megapixel, C-mount, F1.4, includes IR pass filter

Video Performance (H x V Pixels)

  • Default: 1280 x 960 at 30 fps (in combination with 12 mm lens)
  • Optional:
    • 640 x 480 at 60 fps (in combination with 12 mm lens)
    • 800 x 600 at 60 fps (in combination with 8 mm lens)

Maximum frame rates as measured in 96 well plate setup.

Backlight Unit

  • Type: Combined infrared (IR) and white light array
  • Infrared wavelength (typical): 950 nm
  • White light color temperature (typical): 5500 ΒΊK
  • White light intensity range: 0 to Β±10,000 lux (measured directly at the bottom of the water basin)

White light level is controllable via USB (EthoVision XT).

According to published research, zebrafish are sensitive to infrared light, though not above 910 nm. See Shcherbakov et al. 2013, PLoS ONE (5): e64429. The IR light provided by DanioVision is therefore unlikely to affect the fish's behavior.

Note: If you set the Low Light switch to L (low), the highest light level (100%) is around 20 lux.

Note: Light intensity in DVOC-0041 can be set higher than in DVOC-0040. You may want to replicate light conditions set in DanioVision DVOC-0040 in a new experiment with DanioVision DVOC-0041. For example, if you set light intensity to 50% in DanioVision DVOC-0040, this gave 2950 lux. The equivalent percentage to set in DanioVision DVOC-0041 is (2950/5300) x 50 = approximately 28%.

Multi Basin

  • Water inlet, water outlet, and overflow
  • Basin size: 158 x 132 mm
  • Supports ANSI/SBS compatible micro plates (L = 127.76 Β± 0.5 mm (5 1/32" Β± 1/64"), W = 85.48 Β± 0.5 mm (3 23/64" Β± 1/64")) with maximum height of 27 mm (1 1/16")
  • Supports Petri dishes up to 90 mm in diameter
  • Temperature sensor supporting the use of Noldus DVTCU (Temperature Control Unit)

Water Connections

  • Water inlet/outlet: Tube fitting 8 mm
  • Water overflow: Tube fitting 10 mm

Tubes Included

  • 1.5 m Clear PVC Hose, 8 mm OD, 5 mm ID β€” water inlet (green color marked)
  • 1.5 m Clear PVC Hose, 8 mm OD, 5 mm ID β€” water outlet (yellow color marked)
  • 1.5 m Clear PVC Hose, 10 mm OD, 8 mm ID β€” water overflow (red color marked)

PC Interface

  • Interface: USB-2 compatible

External I/O

  • TTL input/output (4x modular RJ45 connector):
    • 4 x 2 TTL input
    • 4 x 2 TTL output
  • TTL expansion port 1:
    • 4 x TTL output (open collector output)
    • 1 x input camera
    • 1 x output camera
  • TTL expansion port 2 (1x modular RJ45 connector):
    • 8-pin feed through

System Start-Up Time

  • Cold start: ~15 minutes

Dimensions

  • Closed position: 61 x 32 x 46 cm (24 1/64" x 12 19/32" x 18 7/64") (h x w x d)
  • Door opening: 36 x 29 cm (14 11/64" x 11 27/64") (h x w)

Weight

  • 15 kg excluding power supply, external tubes, and cables

Cleaning the DanioVision Observation Chamber

We recommend using a proprietary glass-cleaning fluid to clean the outside of the DanioVision Observation Chamber.

Declaration of Conformity

See the Declaration of Conformity documentation included with your unit.

DanioVision Tapping Device

Power

  • Power requirements: 24 Volt DC, 0.4 Amp (max)

Features

  • Controlled from EthoVision XT 10.1 or later with selectable stimulus intensity (1–8)
  • Delivers a discrete 'tap' to the well plate holder
  • Maximum tapping rate of Β±3 taps per second

Connectors

  • Molex 2 pin β€” Power (24V)
  • Molex 4 pin β€” From Tapping Device to control box
  • Multi connector (blue) β€” For TTL control signal

Dimensions

  • Tapping Device (including aluminum bracket): 6 x 24 x 7.5 cm (2 23/64" x 9 29/64" x 2 61/64") (h x w x d)
  • Control box: 2.5 x 10 x 6.5 cm (63/64" x 3 15/16" x 2 9/16") (h x w x d)

Weight

  • ~300 g (Tapping Device, bracket, and control box)

Compatible DanioVision Chambers

  • DVOC-0040
  • DVOC-0041

If not pre-assembled in the DVOC-0041, the package includes mounting material.

DanioVision Toplight Unit

White Light

  • Intensity: >30,000 lux at 25 cm distance (well plate), 1584 lm
  • Color temperature: 4000 K
  • Array: 408 LEDs
  • Max. voltage: 18 Volt

Red Light

  • Luminous flux: 192 lm
  • Wavelength: 623 nm
  • Array: 252 RGB LEDs
  • Max. voltage: 24 Volt

Green Light

  • Luminous flux: 384 lm
  • Wavelength: 525 nm
  • Array: 252 RGB LEDs
  • Max. voltage: 24 Volt

Blue Light

  • Luminous flux: 240 lm
  • Wavelength: 470 nm
  • Array: 252 RGB LEDs
  • Max. voltage: 24 Volt

Unit Dimensions

  • 270 x 220 x 30 mm

Source: DanioVision DVOC-0041 Reference Manual, Noldus Information Technology

Multi-site deployment?

We offer volume licensing and centralized management for labs with multiple locations.

Noldus is here to assist you throughout the whole process.

shopping_bag
check_circle

Thank you!

We'll get back to you shortly.

error

Please correct the following errors:

error

error

error

error

By clicking Submit, you consent to Noldus processing your data as described in our privacy policy.