System Specifications

DanioVision DVOC-0050

Dimensions

  • 61 x 32 x 46 cm (24 1/64" x 12 19/32" x 18 7/64") (H x W x D) in closed position
  • Door opening: 36 x 29 cm (14 11/64" x 11 27/64") (H x W)

Weight

  • 15 kg excluding power supply, external tubes, and cables

Power Supply

  • Power requirements: 24 Volt DC, 4.0 A (max)
  • Mains adapter: input 100–240 Volt AC, output 24 Volt DC, 4.0 A

Camera

  • Brand and type: Basler acA1300-60gm
  • Interface: Gigabit Ethernet (RJ45 connector, CAT5e or higher)
  • Sensor: 1/1.8" Progressive Scan CMOS, monochrome
  • Maximum resolution: 1280 x 1024
  • Maximum frame rate: 60 fps

Lens

  • 12 mm 2 Megapixel, C-mount, F1.4, includes IR pass filter
  • Optional lens: 8 mm 2 Megapixel, C-mount, F1.4, includes IR pass filter. Note: this lens requires a 0.5 mm spacer ring.

Video Performance (H x V Pixels)

  • Default: 1280 x 960 at 30 fps (in combination with 12 mm lens)
  • Optional: 640 x 480 at 60 fps (in combination with 12 mm lens)
  • Optional: 800 x 600 at 60 fps (in combination with 8 mm lens)

Maximum frame rates as measured in a 96 well plate setup.

Backlight Unit

  • Combined infrared (IR) and white light (White Light Stimulus) array
  • Infrared wavelength (typical): 950 nm. According to published research, zebrafish are sensitive to infrared light, though not above 910 nm (see Shcherbakov et al. 2013, PLoS ONE 8(5): e64429). The IR light provided by DanioVision is therefore unlikely to affect the fish's behavior.
  • White light color temperature (typical): 5500 ºK
  • White light level controllable with EthoVision XT. Intensity range: standard levels 0 to approximately 10,000 lux; low light levels (Low Light switch set to L): 0 to about 20 lux. Light level measured directly at the bottom of the water basin.

Multi Basin

  • Water inlet, water outlet, and overflow
  • Basin size: 158 x 132 mm (6 7/32" x 5 13/64")
  • Supports ANSI/SBS compatible micro plates (L = 127.76 ± 0.5 mm, W = 85.48 ± 0.5 mm) with a maximum height of 27 mm (1 1/16")
  • Supports Petri dishes up to 90 mm (3 35/64") in diameter
  • Temperature sensor supporting the use of the Noldus DVTCU Temperature Control Unit

Water Connections

  • Water inlet/outlet: tube fitting 8 mm
  • Water overflow: tube fitting 10 mm

Tubes Included

  • 1.5 m clear PVC hose, 8 mm OD, 5 mm ID for water inlet (green color marked)
  • 1.5 m clear PVC hose, 8 mm OD, 5 mm ID for water outlet (yellow color marked)
  • 1.5 m clear PVC hose, 10 mm OD, 8 mm ID for water overflow (red color marked)

PC Interface

  • Ethernet

External I/O

  • TTL input/output (2x modular RJ45 connector): 2x 2 TTL input, 2x 2 TTL output
  • TTL Trigger Out (2x BNC connector): 2x TTL output

System Start-Up Time (Cold Start)

  • Approximately 15 minutes

Cleaning The DanioVision Observation Chamber

  • Outside of the chamber: use a proprietary glass-cleaning fluid or a detergent solution.
  • Inside parts (Fresnel lens, lens heater, etc.): use only a damp microfiber cloth with a mild detergent solution in warm or lukewarm water. Ensure the cloth is just damp and leaves no droplets on the surface.

DanioVision Temperature Control Unit (TCU)

Dimensions

  • 460 x 210 x 280 mm / 18.11" x 8.26" x 11.02" (L x W x H)
  • Hose connectors: 8 mm outer diameter tube

Weight

  • 7 kg / 15.43 lb

Electrical

  • Operating voltage: 24 V DC
  • Power consumption (average / maximum): 40 / 120 W
  • TTL port specs: 2 input / 2 output, max. 5 V / max. 5 V, open collector

Operational

  • Temperature accuracy: ± 0.5 °C / ± 0.9 °F
  • Temperature set point range: 15–40 °C / 59–104 °F
  • Maximum heating capacity: at least 10 °C / 18 °F above ambient temperature (within the temperature set point range 15–40 °C / 59–104 °F)
  • Maximum cooling capacity: 2.5 °C / 4.5 °F below ambient temperature (within the temperature set point range 15–40 °C / 59–104 °F)
  • Permissible water temperature range: 5–50 °C / 41–122 °F
  • Water capacity: approximately 350 ml / 11.83 oz
  • Pump flow rate – water: approximately 310 ml per min / 10.48 oz per min
  • Water filter type: 20 μm pore size, 1 inch diameter
  • Cooling liquid capacity: approximately 250 ml / 8.45 oz
  • Cooling liquid type: colored, anti-corrosive, anti-algae, PC-type cooling liquid
  • Pump flow rate – cooling liquid: approximately 1650 ml per min / 55.79 oz per min
  • Maximum height difference between DVOC and TCU: 1.3 m / 4.26 ft. Important: do not place the Temperature Control Unit higher than the Observation Chamber.
  • Noise level: average over all frequencies less than 35 dB; maximum peak of approximately 48 dB around 160 Hz

Environmental Conditions

  • Permissible ambient temperature range: operating 10–35 °C / 50–95 °F; storage 5–65 °C / 41–149 °F
  • Permissible relative humidity: 20%–80% (non-condensing)

DanioVision Tapping Device

Dimensions

  • Tapping Device (including aluminum bracket): 6 x 24 x 7.5 cm / 2 23/64" x 9 29/64" x 2 61/64" (H x W x D)
  • Control box: 2.5 x 10 x 6.5 cm / 63/64" x 3 15/16" x 2 9/16" (H x W x D)

Weight

  • Approximately 300 g (Tapping Device, bracket, and control box)
  • If not pre-assembled in the DVOC-0050, the package includes mounting material

Power

  • Power requirements: 24 V DC, 0.4 A (max)

Features

  • Controlled from EthoVision XT 10.1 or later with selectable stimulus intensity (1–8)
  • Delivers a discrete tap to the well plate holder
  • Maximum tapping rate of approximately 3 taps per second

Connectors

  • Molex 2 pin – Power (24 V)
  • Molex 4 pin – From Tapping Device to control box
  • Multi connector (blue) – For TTL control signal

Compatible DanioVision Chambers

  • DVOC-004x, DVOC-0050

DanioVision Toplight Unit

Dimensions

  • 270 x 220 x 30 mm / 10 5/8" x 8 21/32" x 1 3/16"

White Light

  • Greater than 30,000 lux at 25 cm distance (well plate), 1584 lm
  • Color temperature: 4000 °K
  • Array of 408 LEDs
  • Maximum voltage: 18 V

Red Light

  • 192 lm, wavelength 623 nm
  • Array of 252 RGB LEDs
  • Maximum voltage: 24 V

Green Light

  • 384 lm, wavelength 525 nm
  • Array of 252 RGB LEDs
  • Maximum voltage: 24 V

Blue Light

  • 240 lm, wavelength 470 nm
  • Array of 252 RGB LEDs
  • Maximum voltage: 24 V

Source: DanioVision DVOC-0050 Reference Manual - Appendices, Noldus Information Technology

Compare solutions

Not sure which Noldus product fits your protocol? We can help you make the right choice.

Noldus is here to assist you throughout the whole process.

shopping_bag
check_circle

Thank you!

We'll get back to you shortly.

error

Please correct the following errors:

error

error

error

error

By clicking Submit, you consent to Noldus processing your data as described in our privacy policy.

check_circle

Thank you for registering!

We look forward to seeing you at the NoldusViso launch event.

error

Please correct the following errors:

error

error

error

error

error

By clicking Register, you consent to Noldus processing your data as described in our privacy policy.