The DanioVision Tapping Device

Introduction

Larval stage zebrafish display a robust startle response, which is mediated by neural pathways similar to that in higher vertebrates. With the DanioVision Tapping Device you can evoke a startle response in zebrafish larvae.

How the Tapping Device Works

The DanioVision Tapping Device is based on an electromagnetically inductive coil wound around a metal core. The core is made of two parts, a fixed one and a plunger connected to a push pin. When a current flows in the Tapping Device, magnetic force is created between the plunger and the core, causing the plunger and the push pin to quickly move forward. The pushpin hits a metal plate attached to the basin in which the well plate is kept. When deactivated, a spring makes the plunger resume its position.

Controlling the Tapping Device

With the Trial and Hardware Control module of EthoVision XT you can plan when to activate the Tapping Device during a trial. For example, activate the Tapping Device 10 times at intervals of two minutes.

Noldus Test Results: A Summary

  • In both light and dark conditions, the Tapping Device is capable of inducing a startle response in animals of 3 dpf and older.
  • It is in principle possible to design experiments where the animals are subject to stimuli of different intensity. You can set the tap intensity in EthoVision XT.
  • There is no evidence that the startle response measured by EthoVision XT depends on light conditions and location of the subject in the well plate.

Important: We advise you not to compare the results between two DanioVision Observation Chambers provided with the Tapping Device. Slight differences in the position of the Tapping Device may result in differences in the actual stimulus intensity, all else being equal.

Install the DanioVision Tapping Device

If you order the Tapping Device together with a DanioVision Observation Chamber, the Tapping Device is already installed. In all other cases:

  1. Open the lid of the DanioVision Observation Chamber, and remove the bench plate.
  2. Remove the two bolts from the aluminium plate at the bottom of the chamber. Make sure that the backlight box does not move.
    Tip: Use a pencil to mark the exact position of the backlight box on the frame so that you can place it in that position if it accidentally moves.
  3. Put the metal frame of the Tapping Device over the aluminium profiles, and fix the bolts with the supplied spring washers. Use both the already connected spring washers and the new ones.
  4. Connect cables as indicated:
    • The cable with a 4-pin white connector comes from the Tapping Device.
    • The flat cable with a blue connector comes from the bottom of the DanioVision chamber. Raise the metal hooks, insert the connector, then lower the hooks to fix the connector.
    • The cable with a 2-pin connector comes from the bottom of the DanioVision chamber.
  5. Attach the control box to the aluminium profile at the front side of the observation chamber:
    • Insert the screw in one of the holes at the sides of the control box.
    • Attach the spacers and the sliding nuts to the screw.
    • Insert the sliding nuts into the profile, and turn the screw.
  6. Put the bench plate back into place and fix the four screws.

Set Up the Tapping Device in EthoVision XT

Make sure that the DanioVision Observation Chamber is connected to your EthoVision XT computer.

Experiment Settings

  1. Under Trial Control Hardware select Use of Trial Control Hardware, then click Settings.
  2. In the Hardware Interface window, select DanioVision DVOC-0050/TPC.
    Tip: You can find your DanioVision version number on the label at the back of the chamber.
  3. When needed, adjust the default light settings for the White Bottom Light. Make sure the message at the bottom of the Hardware Interface window reads Connected. Click Refresh when necessary.
  4. Click OK and go to the next section.

Arena Settings

Note: If you created an experiment with a pre-defined template and selected an arena template for DanioVision with the Tapping Device (DanioVision DVOC 00xx/T, xx wells), the Tapping Stimulus has been assigned automatically to the arenas. You can skip the steps below and go to the next section.

  1. Open the Arena Settings you want to use.
  2. In the Arena Settings pane click Arena - Hardware Mapping and then click the Arena - Hardware Mapping button.
  3. Click Add Device.
  4. In the new row that appears, under Device Type, select Tapping Stimulus.
  5. Under A1 or Arena 1 (depending on how the arenas are named), select Tapping Stimulus.
  6. To test the Tapping Device, select the cell under A1/Arena 1 and click Test.

Trial Control Settings

  1. Open the Trial Control Settings containing the protocol in which you want to insert the Tapping Stimulus. Under Actions - Hardware click the button next to Tapping Stimulus.
  2. In the Hardware Action window:
    • Enter the Action Name.
    • Specify the Intensity (0% to 100%).
    • Click Test to get an impression of the tap intensity.
    • When ready, click OK.
  3. Insert the Action box in your Trial Control protocol.
  4. Insert Condition boxes and, when required, Sub-rules to create routines. Test the protocol before carrying out the real trials.

Important: Allow at least 300 milliseconds between two consecutive tapping events. To do so, place a Time Condition box that specifies this time (> 0.3 sec) between two Tapping Stimulus Action boxes.

Note: If the Tapping Device does not respond to the test action, for example after it has not been in use for a long time, open the bench plate and use a screwdriver to gently push the pushpin back, then release it so that it hits the metal plate, as if it was activated. Do this a few times and then test the action again. Also check the connections inside.

Tapping Stimulus Routine

If you have created an experiment from a DanioVision template with the Tapping Stimulus Device (DVOC-0050_T), you can find an example of a routine with the Tapping Stimulus under Trial Control Settings.

Analyze the Response to the Tapping Stimulus

Aim

You can analyze the startle response by looking at dependent variables like Activity, Distance Moved, or Acceleration (state) immediately before and after the administration of the stimulus.

Example

We examine the Total Distance Moved in two intervals: (1) a 5-second interval immediately before the tapping stimulus command, used as a baseline, and (2) a 1-second interval after the tapping stimulus command. We expect that the total distance moved increased from "before" (assumed to be a baseline) to "after".

Procedure

  1. Open or create a Data Profile.
  2. Interval "before": In the Data Profile, under Nesting click the button next to Free Interval. Define an interval from 5 seconds before the action "Tapping Stimulus" was active to 0 seconds after the action "Tapping Stimulus" was active.
  3. Interval "after": In the Data Profile, under Nesting click the button next to Free Interval. Define an interval from 0 seconds before the action "Tapping Stimulus" was active to 1 second after the action "Tapping Stimulus" was active.
  4. Click the button under Common Elements, next to Result. Place the Result box under the existing one. Each of the resulting Nest boxes is connected to a separate Results box.
  5. In the Analysis Profile define Distance Moved.
  6. Choose Analysis > Results > Statistics and Charts.

Source: DanioVision DVOC-0050 Reference Manual - Tapping Device, Noldus Information Technology

Request a trial license

Try the software in your own lab before you buy. We offer trial periods for qualifying researchers.

Noldus is here to assist you throughout the whole process.

shopping_bag
check_circle

Thank you!

We'll get back to you shortly.

error

Please correct the following errors:

error

error

error

error

By clicking Submit, you consent to Noldus processing your data as described in our privacy policy.