DanioVision DVOC-0050 - Temperature Control Unit - Feedback and Error Messages

Last updated: Jul 28, 2026

Feedback and Error Messages

The DanioVision Temperature Control Unit gives feedback about its status with four LED lights and error codes on the temperature display.

LED Indicators

  • Filling/Draining – Orange blinking: System is pumping. The system can be filled with water, water can be drained from the basin, or cooling liquid can be filled.
  • Filling/Draining – Red continuous: Time out. There has been no flow in the water circulation and/or the cooling liquid circulation for more than 10 minutes. Reset the system by turning the Operation switch to Off and then back to Fill/Drain.
  • Heating/Cooling – Green blinking: The water is being heated or cooled to the set temperature. Wait; no action needed.
  • Ready – Green continuous: The set temperature has been reached and the system is ready to be used. An experiment can be started.
  • Error – Red, 1 blink repeatedly: Insufficient flow in the water circulation. Look for the source of the problem. This may be:
    • Low water level in the basin due to evaporation or leakage. In this case, fill the system.
    • Flow blockage somewhere in the system or filter clogged. Replace the filter or contact Noldus.
    • Pump failure.
    The Fill/Drain mode can be used to localize the problem. Reset the system by setting the Operation switch to Off and then to Fill/Drain.
  • Error – Red, 2 blinks repeatedly: Insufficient flow in the cooling liquid circulation. Look for the source of the problem. This may be:
    • Low cooling liquid level in the pump reservoir. Check if the liquid level is above the indication. If not, refill cooling liquid and check for signs of leakage in the emergency liquid container.
    • Flow blockage somewhere in the system.
    • Pump failure.
    The Fill/Drain mode can be used to localize the problem. Reset the system by setting the Operation switch to Off and then to Fill/Drain.
  • Error – Red, 3 blinks repeatedly: The fan is not spinning, or not spinning at sufficient speed. Check if there is anything blocking the fan. If this is not the case, please contact Noldus support.
  • Error – Red, 4 blinks repeatedly: System is overheated. Turn off the system and end the experiment. Cooling liquid can be cooled down by switching the Operation switch to Fill/Drain. Search for the possible cause of the overheating:
    • Air vents of the Temperature Control Unit might be blocked so the radiator cannot dissipate heat. Make sure air vents are free.
    • System is used in environmental conditions that have not been tested, for example, room temperature above 35°C.
  • Error – Red, 5 blinks repeatedly: Heating and cooling have been activated simultaneously. An electrical problem has occurred. Please contact Noldus support.
  • Error – Red, 6 blinks repeatedly: A problem with the data logger has occurred. Open the TCU and take the SD card out.
    • If it is locked, unlock it.
    • Check that it has a minimum capacity of 4 GB.
    Put the SD card back in the TCU. If the error persists, take the card out and format it (FAT32). Leave the card empty; do not copy the old files to the newly formatted card. Put the SD card back in the TCU and restart the TCU.
  • Error – Red continuous: A temperature deviation larger than 0.5°C above or below the set temperature has been detected. Water conditioning will continue to try to resolve the situation. Once resolved, the status returns to 'Ready'. Depending on your protocol you can choose to continue or abort the experiment.

Temperature Display Codes

  • Code 1: The water is being heated to the set temperature.
  • Code 2: The water is being cooled to the set temperature.
  • Code 3: The current water temperature is at least 0.5°C below the set temperature. It is shown together with temperature alarm code E05. In normal conditions, when code 3 is shown, code 1 is also shown and the red Error LED (continuous) is on.
  • Code 4: The current water temperature is at least 0.5°C above the set temperature. It is shown together with temperature alarm code E04. In normal conditions, when code 4 is shown, code 2 is also shown and the red Error LED (continuous) is on.
  • Error mode: This is a system setting indicator, which blinks when the TCU is in heating/cooling mode. No action need be taken.

Error Codes

  • LCD is off: A possible reason is that one of the flat cables coming from the LCD display is not properly connected. Open the DV-TCU, locate the multi-color flat cable (3rd to the right of the SD card), disconnect it and connect it again.
  • E01: The temperature sensor is not detected. Please connect the temperature sensor cable (with cinch-plugs on both ends).
  • E04: 'Too high' temperature alarm. The measured temperature is 0.5°C higher than the set temperature. Depending on your protocol you can choose to continue or abort the experiment. Please note that this error also occurs while the system is heating/cooling before the start of the experiment; the system continues to heat/cool to try to resolve the error.

Error Codes

Error Code Explanation What to Do
E05 'Too low' temperature alarm. The measured temperature is 0.5 °C lower than the set temperature. Depending on your protocol you can choose to continue or abort the experiment. Please note that this error also occurs while the system is heating/cooling before the start of the experiment.
Ed1 System error. See the Status = Error section for further guidance.
nO Another display has been selected accidentally.
  1. Press the up or down arrow until the code b1 appears on the display.
  2. Press the Set button for three seconds to confirm.

Source: DanioVision DVOC-0050 Reference Manual - Temperature Control Unit, Noldus Information Technology

Planning a new study?

We can advise on the optimal hardware and software configuration for your upcoming research project.

Noldus is here to assist you throughout the whole process.

shopping_bag
check_circle

Thank you!

We'll get back to you shortly.

error

Please correct the following errors:

error

error

error

error

By clicking Submit, you consent to Noldus processing your data as described in our privacy policy.