Track3D Calibration Error Caused by Reflective Spots in the Calibration Image
Last updated: Nov 26, 2026
Applies to: Track3D 2.0 (reported in one ticket; the other ticket does not state a version)
Problem
An error appears in the DOS command window (for example 'Error while evaluating uicontrol Callback') when you create a calibration file after moving a camera or setting up a new calibration.
Details
Two tickets. In both, Support found too many white reflective spots in the calibration image. These interfere with the recognition of the calibration markers in the calibration frame. Removing the spots in an image editor (Paint) solved the problem. One ticket mentions Track 3D 2.0, the other does not state a version. The ticket names the calibration program CenterFind; the Track3D manual calls it CentroidFinder. The Track3D manual describes the same fix under Edit the Calibration Images.
Solution
- Open the calibration image that was used for the calibration file (with two camera views, both calibration images) in an image editor such as Paint.
- Remove the white or bright spots that are not calibration markers, especially those close to a marker, so that they are no longer bright.
- Save the image under a new name as a BMP file with a bit depth of 24.
- Repeat for the second calibration image if you use two camera views, and calibrate again in CentroidFinder with the new images.
- If the error continues, contact Noldus Support.